首頁 > 產(chǎn)品展示 > > 其他檢測儀器 > TC-500 氧氮聯(lián)合測定儀

TC-500 氧氮聯(lián)合測定儀

簡要描述:簡單說明:TC500系列儀器質(zhì)量、性能、工藝滿足您對金屬、難熔物、其他無機材料日常生產(chǎn)中氮/氧分析的需要

  • 產(chǎn)品型號:TC-500
  • 廠商性質(zhì):生產(chǎn)廠家
  • 更新時間:2024-10-08
  • 訪  問  量:1747
詳情介紹

詳細說明:

TC500系列儀器質(zhì)量、性能、工藝滿足您對金屬、難熔物、其他無機材料日常生產(chǎn)中氮/氧分析的需要

Determination of Oxygen and Nitrogen in
Reactive/Refractory Metals and Their Alloys*
 
Approval
ASTM E-1409 Oxygen Determination in Titanium and Titanium Alloys
ASTM E-1569 Oxygen Determination in Tantalum
ASTM E-1937 Nitrogen Determination in Titanium and Titanium Alloys
 
Sample Preparation
A clean representative sample is required in order to obtain optimum results. Solid samples can be leached in suitable acid or abraded with a clean file, rinsed in acetone, and dried with warm air. Refer to ASTM methods above for further details.
 
Accessories
782-720 Crucible; 782-721 Electrode Tip; 502-344 UHP Nickel
Baskets; 501-073 Graphite; 503-032 Glass Accelerator Scoop;
501-059 Tin Capsules—for powdered samples. (Additional steps required for powdered samples are noted in parenthesis.)
 
Calibration Standard
LECO 501-320 Titanium Pin, 502-047 Zirconium Pin, NIST, or other suitable reactive/refractory metal standard.
 
Method Parameters
Analysis Parameters
Outgas Cycles                           2
Analysis Delay (seconds)             20
Analysis Comparator                 1
Analysis Type                              Auto
 
Element Parameters                  Oxygen           Nitrogen
Minimum Time (seconds)
35
55
Integration Delay (seconds)
5
15
Comparator Level (%)
1
1
 
Furnace Parameters
Furnace Control Mode             Power
Purge Time (seconds)                    15
Outgas Time (seconds)                20
Cool Time (seconds)                    5
Outgas Power (Watts)                 6300
Analyze Power (Watts)                5300
 
 
 
 
 


 

*This includes Ti, Zr, W, Mo, Ta, Nb, Hf, and their alloys.

 
TC500


 

 
Typical Results
 
 
 
Titanium Pin
Mass (g)
Oxygen (%)
Nitrogen (%)
 
0.1135
0.1808
0.0189
 
0.1133
0.1843
0.0194
 
0.1134
0.1831
0.0184
 
0.1138
0.1819
0.0188
 
0.1133
0.1821
0.0195
 
0.1131
0.1829
0.0185
 
0.1134
0.1833
0.0191
 
0.1143
0.1849
0.0194
 
0.1125
0.1837
0.0189
 
0.1072
0.1830
0.0191
 
Average
0.1830
0.0190
 
Std. Dev.
0.0012
0.0004
 
Zirconium Wire
 
Mass (g)
 
Oxygen (%)
 
Nitrogen (%)
 
0.1067
0.1246
0.0017
 
0.0978
0.1273
0.0017
 
0.1180
0.1246
0.0019
 
0.1097
0.1235
0.0014
 
0.1141
0.1272
0.0017
 
0.1017
0.1226
0.0018
 
0.1140
0.1269
0.0020
 
0.0987
0.1239
0.0020
 
0.1015
0.1233
0.0022
 
0.1197
0.1270
0.0020
 
Average
0.1251
0.0018
 
Std. Dev.
0.0018
0.0002
 
Procedure
 
 
 
1.   Determine the blank as follows:
a.   Enter the "blank" ID code with a 1.0000 g weight in the weight stack.
b.   Press the loader control switch, the sample loader will open.
c.   Place one 502-344 UHP Nickel basket into the loading head using clean tweezers.
(Place a 501-059 Tin capsule into the Nickel basket before placing it in the loading head.)
d.   Press the loader control switch, the sample loader will close and seal and the furnace electrode will open.
e.   Remove crucible from electrode tip and discard.  Clean furnace area using the appropriate brushes.  Vacuum away loose dust.
f.    Place ~0.05 g 501-073 into the bottom of a 782-720 crucible.  ~0.05 g is approximay a ¼ full
503-032 glass accelerator scoop.
g.   Place the crucible on the lower electrode.
h.   Press the loader control switch, the furnace electrode will close and the analysis sequence will start automatically.
i.    Repeat steps 1a through 1h at least four more times.
j.    Enter blank following routine outlined in operator's instruction manual.
2.   Calibrate the instrument as follows:
a.   Weigh the calibration sample. (Weigh the calibration sample into the tin capsule.)
b.   Enter the calibration sample ID code and sample weight in the weight stack.
c.   Place the calibration sample (capsule) into a nickel basket.
d.   Press the loader control switch, the sample loader will open.
e.   Carefully place the calibration sample/nickel basket (capsule) into the loading head using clean tweezers.  Make sure that the calibration sample (capsule) stays in the basket and the basket
stays upright.
f.    Press the loader control switch, the sample loader will close and seal and the furnace electrode will open.
continued on page 3


 

Procedure (continued from page 2)
g.   Remove crucible from electrode tip and discard.  Clean furnace area using the appropriate brushes.  Vacuum away loose dust.
h.   Place ~0.05 g graphite into the bottom of a crucible.
i.    Place the crucible on the lower electrode.
j.    Press the loader control switch, the furnace electrode will close and the analysis sequence will start automatically.
k.   Repeat steps 2a through 2j a minimum of three times.
l.    Complete a calibration by following the auto calibration procedure as outlined in the operator's instruction manual.
m.  Verify the calibration by analyzing the calibration sample again.  It should fall within the expected tolerances.  If not, repeat steps 2a through 2l again.
3.   Analyze the samples as follows:
a.   Weigh ~0.1 g sample. (Weigh the sample into the tin capsule.)
b.   Enter the sample ID code and sample weight in the weight stack.
c.   Place the sample (capsule) into a nickel basket.
d.   Press the loader control switch, the sample loader will open.
e.   Carefully place the sample/nickel basket (capsule) into the loading head using clean tweezers.
Make sure that the sample (capsule) stays in the basket and the basket stays upright.
f.    Press the loader control switch, the sample loader will close and seal and the furnace electrode will open.
g.   Remove crucible from electrode tip and discard.  Clean furnace area using the appropriate brushes.  Vacuum away loose dust.
h.   Place ~0.05 g graphite into the bottom of a crucible.
i.    Place the crucible on the lower electrode.
j.    Press the loader control switch, the furnace electrode will close and the analysis sequence will start automatically.
 
Theory of Operation
The TC500 is a microprocessor-based, software-controlled instrument that measures both nitrogen and oxygen in a wide variety of metals, refractories, and inorganic materials. The inert gas fusion principle is employed. A weighed sample, placed in a high-purity graphite crucible, is fused under a flowing helium gas stream at temperatures sufficient to release oxygen, nitrogen, and hydrogen. The oxygen in the
sample, combines with the carbon from the crucible to form carbon monoxide. The nitrogen present in
the sample releases as molecular nitrogen, and any hydrogen present is released as hydrogen gas. The helium carries the sample gases through heated rare earth copper oxide which converts carbon
monoxide to carbon dioxide (CO2) and hydrogen to water (H2O). The nitrogen passes through
unchanged. The gases are then passed through a CO2  infrared (IR) cell where the oxygen is measured as
CO2. CO2  and H2O are then removed by a Lecosorb/Anhydrone trap, while the nitrogen passes through
to a thermal conductivity (TC) cell for determination.


 

TC500 Flow Diagram
 
 
 
 
 

留言框

  • 產(chǎn)品:

  • 您的單位:

  • 您的姓名:

  • 聯(lián)系電話:

  • 常用郵箱:

  • 省份:

  • 詳細地址:

  • 補充說明:

  • 驗證碼:

    請輸入計算結(jié)果(填寫阿拉伯?dāng)?shù)字),如:三加四=7
技術(shù)支持:化工儀器網(wǎng)   sitemap.xml   管理登陸
©2025 版權(quán)所有:寧波市鄞州瑾瑞儀器設(shè)備有限公司   備案號:浙ICP備14028217號-3

浙公網(wǎng)安備33020402000311號

日本肥老妇色xxxxx日本老妇| 97亚洲狠狠色综合久久久久| 西瓜视频高清在线观看免费| 曰韩人妻无码一区二区三区综合部| 中国帅气体育生gary视频| 女人被强╳到高潮喷水在线观看| 成av人电影在线观看| 国产精品久久久久jk制服| 青丝影院视频在线播放 | 亚洲av片一区二区三区| 精品久久久久成人码免费动漫| 上司人妻互换中文字幕| 欧美三级在线播放| 黄瓜成视频人app| 日日碰狠狠躁久久躁9| 久久强奷乱码老熟女网站| 在线无码VA中文字幕无码| 国产精品无码AV天天爽播放器 | 撸啊撸视频在线观看| 国产成人精品a视频| 国产伦精品一区二区三区妓女| 国产盗摄xxxx视频xxxx| 无码观看aaaaaaaa片| 试看20分钟做受视频| 国产亚洲精品精华液| 麻豆亚洲av熟女国产一区二| 女女互揉吃奶揉到高潮视频| 久久久无码人妻精品无码| 久久午夜夜伦鲁鲁片无码免费| 99久久国产福利自产拍| 高清中文字幕在线a片| 欧美噜噜久久久XXX| 色哟哟网站在线观看| 99久久久无码国产精品| 亚洲精品一区三区三区在线观看 | 日本少妇高潮喷水xxxxxxx| 国产在线精品二区| 欧美成人精品第一区二区三区| 人妻少妇精品无码专区动漫| 日本熟妇人妻中出| 亚洲欧洲∨国产一区二区三区|